Floor 1, No. 3, Shugang Avenue, Hongmei Town, Dongguan City, Guangdong Province, China

20 years environmental test chamber manufacturer3000+ customer deliverables worldwide           Email: shirley@deruitest.com
WhatsApp Consultation:+86 135 2795 8363

DR-H305 Ozone Aging Test Chamber for Rubber

The DERUI DR-H305 ozone aging test chamber exposes rubber specimens to controlled ozone and temperature under method-defined static or dynamic tensile strain. Select the ozone range, fixture and environmental functions for the specified test procedure.

Model DR-H305
Test type Static or dynamic tensile ozone exposure
Ozone range 10–1000 pphm or 1–500 ppm; select the ordered option
Specimens Vulcanized or thermoplastic rubber, seals and cable materials
Before quotation Confirm the method, ozone setpoint, strain programme and specimen layout
Safe and reliable
well-made
Quality assurance
customization
MessageWhatsApp
Hotline number 15580327593
  • Product Details
  • Technical parameters
  • FAQ
  • Contact us
DR-H305 ozone aging test chamber

Control Ozone, Strain and Exposure Conditions Around the Rubber Test Method

Ozone cracking is evaluated under defined conditions, not by chamber concentration alone. The test plan must connect the ozone level with specimen strain, temperature, air movement, exposure time and the crack-evaluation method.

The DR-H305 is offered for laboratories comparing rubber compounds, molded or extruded rubber parts, seals, hoses, cable insulation and sheathing. The ordered configuration can support static or dynamic tensile fixtures. Final suitability depends on the cited standard, edition and specimen geometry.

Match the Chamber to the Ozone-Cracking Procedure

ISO 1431-1:2024 addresses static and dynamic tensile-strain exposure of vulcanized or thermoplastic rubber in ozone-containing air at a defined temperature and, when required, relative humidity. ASTM D1149-18(2025) also evaluates cracking in an ozone-controlled environment under static or dynamic strain. Neither method turns one generic chamber programme into a universal product-life prediction.

Buyer requirement DR-H305 configuration direction Confirm before quotation
Static-strain ozone cracking Static tensile holders and rotating rack where required Specimen shape, extension, holder count, orientation and inspection interval
Dynamic-strain ozone cracking Dynamic tensile mechanism with method-specific motion Strain limits, frequency, cycle pattern, loaded quantity and fixture drawing
Cable or customer-specific procedure Reviewed fixture, ozone range and environmental controls Exact document, edition, clause, specimen preparation and acceptance criteria

Use the current ISO 1431-1 or applicable customer method to define the procedure. ISO 1431-3 covers reference and alternative methods for determining ozone concentration in laboratory chambers.

Ozone Generation, Measurement and Chamber Control

Ozone-generation range

The published DR-H305 options cover 10–1000 pphm for low-concentration work or 1–500 ppm for high-concentration work. Select the required option at ordering; these ranges are not interchangeable defaults.

Concentration measurement

Define the analyser principle, sampling location, display unit, calibration method and required traceability. Confirm how the system records stabilization, exposure and alarm events.

Air movement and exchange

The current specification lists 12–16 mm/s chamber air velocity. Confirm the measurement location, loaded condition, exchange requirement and distribution evidence for the governing procedure.

Temperature and humidity

Choose the published temperature and humidity option only when the method requires it. The quotation should state the operating range and control performance for the selected ozone condition.

Light exclusion

Ozone-cracking procedures exclude direct-light effects. Confirm chamber construction, viewing arrangement and inspection practice maintain the required exposure environment.

Ozone destruction and exhaust

Define the destruction or exhaust path, interlocks and site connection before installation. Ozone is hazardous and must not be released into the work area.

Select the Fixture From the Specimen and Strain Programme

Fixture choice affects exposure validity and usable capacity. Provide a specimen drawing rather than estimating capacity from chamber volume. DERUI can review the clamping method, extension, clearance and number of simultaneously exposed pieces.

Static fixture input
Specimen type, gauge section, required extension, clamping area, orientation, quantity and rack rotation requirement.
Dynamic fixture input
Minimum and maximum strain, frequency, waveform or motion sequence, cycle duration, quantity and stop conditions.
Evaluation input
Visual grading or image analysis, inspection intervals, crack criterion, magnification, conditioning and control specimens.

The published rotating sample rack is listed at 1 r/min. Dynamic tensile movement is a separate fixture function and must be defined from the procedure; rack rotation alone does not establish dynamic-strain capability.

Illustrative view of DERUI ozone aging chamber specimen rack with rubber test strips

Monitoring, Alarms and Test Records

A useful record shows that the loaded test remained within the agreed conditions. Define the channels, sampling interval and export format in the purchase specification.

Record or function Buyer decision Acceptance evidence
Ozone trend Unit, setpoint, tolerance, interval and retention Loaded-run trend and sample export
Environmental conditions Required temperature, humidity and air parameters Sensor identification and recorded values
Fixture operation Rack speed or dynamic motion and duration Demonstration with representative specimens
Fault response High ozone, loss of flow, door opening and power interruption Alarm list, interlock test and recovery behavior

What to Send for a Model-Specific DR-H305 Proposal

A quotation can only be checked against the intended procedure when the test inputs are complete. Send these items together:

Method
Standard, edition, clause, ozone unit, concentration, temperature, humidity and duration.
Specimens
Material, dimensions, shape, quantity, preparation, conditioning and exposed surfaces.
Strain programme
Static extension or dynamic limits, frequency, cycle sequence and holder arrangement.
Evaluation
Inspection times, crack grading, image analysis, magnification and acceptance criteria.
Records
Trend channels, export format, calibration documents, user access and alarms.
Installation
Voltage, room conditions, exhaust route, access, destination requirements and training.
Configure the chamber from the ozone test methodSend the procedure and specimen drawing for fixture and control review.

Request a DR-H305 Proposal

Safety, Factory Review and Delivery Acceptance

Before shipment, agree the test load and evidence used to verify the ordered scope. A useful factory review covers concentration control, environmental conditions, fixture movement, records and safety response.

  1. Approve the chamber, fixture and installation drawings.
  2. Confirm the ozone analyser, unit, calibration scope and sampling location.
  3. Demonstrate stabilization and control at agreed representative conditions.
  4. Check rack rotation or dynamic fixture motion with a representative load.
  5. Test alarms, door interlock, ozone shutdown and destruction or exhaust response.
  6. Review trend exports, manuals, maintenance tasks, consumables and spare parts.

Ozone is hazardous. The laboratory remains responsible for site ventilation, exposure controls, operating procedures and applicable safety regulations. Equipment verification does not replace specimen testing and evaluation under the approved method.

Request an Ozone-Chamber Configuration Review

Send the method, ozone concentration, specimen drawing, strain programme, evaluation intervals and site utilities. DERUI can prepare the DR-H305 technical offer around those inputs.

Discuss My Ozone Test

FacebookLinkedInXRedditWhatsApp

Published DR-H305 Technical Configuration

The values below consolidate the configuration currently published for this product. Select the low- or high-concentration ozone system and the required environmental and fixture options when ordering. The signed technical agreement and quotation control the delivered specification.

Item Published configuration
Model DR-H305
Temperature range Option A: 0–70°C; Option B: RT+10–70°C. Specify at ordering.
Humidity range Option A: 40–98% RH; Option B: 85–98% RH. Confirm whether humidity control is required by the selected method.
Temperature fluctuation / uniformity / deviation ≤±0.5°C / ≤±2°C / ≤±1.5°C
Humidity fluctuation +2%, −3% RH
Heating rate 0.7–1.0°C/min
Published cooling statement Room temperature to 10°C in ≤20 min / non-limiting; confirm the ordered option and test condition.
Ozone concentration options 10–1000 pphm (low concentration) or 1–500 ppm (high concentration). Specify at ordering.
Published chamber air velocity 12–16 mm/s
Sample rack Static or dynamic tensile test rack; fixture selected from the specimen and procedure
Rack rotation 360° rotating sample rack; published speed 1 r/min
Timer 0–9999 h
Power supply AC 380 V ±10%, 50 Hz, three-phase five-wire; destination supply must be confirmed
Standards planning Configuration review for ISO 1431-1, ISO 1431-3, ASTM D1149 and applicable cable or customer procedures. Suitability depends on the exact edition and ordered functions.

Ozone Aging Test Chamber FAQs

What materials can the DR-H305 evaluate?

Typical specimens include vulcanized or thermoplastic rubber, elastomeric seals, hoses, molded or extruded rubber, and cable insulation or sheathing. The governing product method determines specimen preparation and acceptance.

What is the difference between static and dynamic ozone testing?

Static testing holds specimens at a defined tensile strain during exposure. Dynamic testing repeatedly changes strain according to a defined motion and frequency. They require different fixture capability and must not be treated as the same rack setup.

Which ozone concentration range should I select?

Select the range from the exact method and test level. The current product lists 10–1000 pphm and 1–500 ppm as separate options. Send the required unit, setpoint and tolerance so DERUI can confirm the analyser and generator configuration.

Does the chamber automatically comply with ISO 1431-1 or ASTM D1149?

No equipment name alone proves method compliance. Confirm the standard edition, ozone measurement, environmental conditions, air movement, fixture, specimen plan and records against the ordered configuration.

Can ozone-test hours predict outdoor service life?

Not by a universal conversion. Accelerated ozone exposure compares resistance under defined laboratory conditions. Correlation with service depends on material, strain, temperature, ozone environment and the failure criterion.

What safety provisions should be reviewed?

Review door interlocks, high-ozone alarms, shutdown logic, ozone destruction or exhaust, leak management, room ventilation and maintenance access. Site risk controls and regulatory compliance remain the laboratory's responsibility.

What should I send for a quotation?

Send the standard and edition, concentration, temperature and humidity, specimen drawing and quantity, static or dynamic strain programme, inspection plan, data requirements, destination voltage and exhaust conditions.

WhatsAppWhatsApp
WhatsAppWhatsApp
WhatsApp
MessageMessage

Request a Quote