DR-H502-2
IPX3 IPX4 Rain Spray Test Chamber DR-H502-2
The DERUI DR-H502-2 uses an oscillating tube and rotating specimen support for spraying and splashing tests on electrical enclosures. Choose the tube and mounting arrangement around the complete assembly, not enclosure volume alone. A specimen-fit review checks moving clearance, exposed surfaces and the water-delivery verification required by your procedure.
DR-H502-2
Choose the Spray Geometry Your Enclosure Needs
Repeatable spraying and splashing—not immersion
DR-H502-2 is intended for enclosure testing where the specified exposure is delivered by an oscillating tube. Applications can include electrical housings, outdoor instruments, lighting assemblies and components with seals or connections that need controlled water exposure.
The moving tube distributes water around the specimen. Table rotation changes the presented surfaces, and height adjustment helps position the assembly relative to the spray geometry. These mechanisms make mounting and coverage part of the test setup—not an afterthought.
Use the chamber to compare sealing designs during development and to execute a documented qualification procedure. Record the fixture and operating state as well as the water exposure so later tests can reproduce the same arrangement.
Start with the complete assembly: a cable gland or mounting bracket can change both the part’s clearance and the path water takes toward its seals.

DR-H502-2
IPX3 and IPX4: Coverage Is More Than an Angle
| Setup question | IPX3 spraying water | IPX4 splashing water |
|---|---|---|
| Exposure intent | Spraying water within the prescribed angular region around the vertical | Splashing water from the directions specified by the method |
| Spray arrangement | Use the prescribed active-hole region and oscillating-tube setup | Use the prescribed broader spray arrangement and specimen support |
| Motion | Verify the selected sweep, speed/cycle and exposure time | Verify the required sweep, speed/cycle and surface coverage |
| Specimen positioning | Place the assembly at the required reference position without obstructing the spray | Check support, orientation and whether all required surfaces receive exposure |
| Evaluation | Apply the product-specific harmful-ingress and functional/electrical criteria | Apply the product-specific harmful-ingress and functional/electrical criteria |
IEC 60529 provides the enclosure IP-code framework. Its IPX3/IPX4 procedures distinguish the oscillating-tube arrangement from the spray-nozzle arrangement. Identify which apparatus your product standard or customer procedure requires.
DR-H502-2
DR-H502-2 Technical Parameters
| Parameter | Configuration |
|---|---|
| Model | DR-H502-2 |
| Test arrangement | Oscillating-tube spraying and splashing water for IPX3 / IPX4 applications |
| Internal workspace (L × D × H) | 900 × 900 × 900 mm; nominal geometric volume 729 L |
| External dimensions (L × D × H) | 1400 × 1120 × 1750 mm |
| Oscillating-tube radius | R400 mm; R200 mm listed as an option |
| Listed tube travel | 120° for IPX3; 350° for IPX4; custom settings available. Verify the travel definition and method-specific motion during acceptance |
| Specimen-table motion | Continuous forward rotation or alternating forward/reverse rotation |
| Specimen-table height adjustment | 350–600 mm |
| Listed table load | 50 kg; include specimen and fixture in the loading review |
| Equipment supply / rated power | 220 V / 1.0 kW; destination frequency and connection defined in the proposal |
| Listed specimen supply | 220 V; electrical interface, allowable load and wet-test protection defined by the selected arrangement |
| Protection functions | Electrical leakage protection, water-shortage protection and short-circuit protection |
| Listed warranty | 1 year; coverage and service terms specified in the quotation |
The 729 L figure is calculated from cabinet dimensions; usable specimen capacity is governed by the tube, support and clearances. For a method-matched configuration, agree spray-hole geometry, water delivery, motion cycle, table dimensions and timing requirements alongside these mechanical specifications.
DR-H502-2
Will Your Assembly Fit the R400 Spray Area?
Check the swept envelope
Include the fixture, cable bends, protruding connectors and the changing envelope during rotation. Keep the assembly clear of the moving tube across the complete programmed travel.
Align height and support
The listed table height is adjustable from 350 to 600 mm. Check the relevant specimen reference point against the tube center and use a support that follows the method without masking surfaces.
Account for fixture mass
Use the listed 50 kg table load for preliminary selection, then review combined specimen/fixture mass, mounting stability and off-center loading. Do not size the setup by specimen mass alone.
Fixture choices can change the result
A plate placed beneath the enclosure may shield the underside; a clamp may compress a gasket; an incorrectly routed cable may divert water away from a gland. The fixture should hold the specimen in the required orientation without improving its sealing performance or hiding an exposure surface.
If you want to test multiple units, submit a layout showing spacing, mounting and water access to each specimen. More parts fitting in the cabinet does not establish that each receives an acceptable exposure.
DR-H502-2
Water Delivery, Test Records and Factory Acceptance
Verify flow for the installed tube
Required total flow depends on the selected tube and active spray holes. Check the applicable method’s per-hole delivery and open-hole arrangement; total flow follows Q = N × q when N active holes each deliver q. Do not apply the same total flow to R200 and R400 without checking their geometry.
Before testing, inspect for blocked holes, confirm the flow indication and check tube motion and specimen position. A pressure reading alone cannot establish the correct water delivery. Follow the supplied water-quality and cleaning instructions to limit deposits and maintain the spray pattern.
| Acceptance item | What to agree and demonstrate |
|---|---|
| Tube and water distribution | Installed radius, hole geometry and active-hole arrangement for each required exposure |
| Flow measurement | Specified operating flow, instrument calibration and verification procedure |
| Motion and timing | Angle definition, travel, cycle/speed and exposure timing at the agreed settings |
| Specimen setup | Table diameter, usable height, loaded rotation, fixture clearance and required coverage |
| Electrical operation | Equipment supply, sample supply/load, cable routing and protection checks |
| Records and documents | Setup drawing, measured settings, test record format, manual, maintenance and agreed acceptance report |
Separate exposure records from product approval
Record specimen ID, sealing configuration, mounting, test settings, water conditions, exposure duration and any interruptions. After testing, use the prescribed handling and inspection sequence; assess harmful ingress, operation and any electrical checks against the governing document.
Automatic logging, USB export, remote monitoring and water-temperature control should be specified separately if your laboratory needs them. For site planning, provide voltage/frequency, water supply, drain location, access route and service clearance. Maintain spray holes and filters where fitted, and never bypass wet-test electrical protection.
DR-H502-2
Compare the complete test setup before ordering
For purchasing
Compare the installed tube, active spray holes, total water delivery and full loaded motion. Include specimen-table dimensions and the rotating fixture, not just the 900mm cabinet. Ask each supplier to demonstrate the selected angle convention and motion recipe against your adopted procedure. The listed 120° and 350° settings require a method-matched acceptance check rather than an automatic compliance claim.
For the application engineer
For a housing with an offset cable gland, show the cable route and its swept envelope while the table turns. Check that clamps neither shelter the gland nor collide with the tube. For a batch, show every enclosure and support within the useful spray area; the number that fits on a shelf is not the number receiving an acceptable exposure.
Evidence to include
Agree measurements for the installed tube, hole arrangement, flow, motion and timing. Demonstrate the required sequence with the representative fixture and record any shielding or collision issue. Pressure indication alone is not a substitute for verifying the water delivered by the active holes.
DR-H502-2
Options to specify in the configuration proposal
Request interchangeable fixtures, a reviewed alternative tube, specimen monitoring and digital test records where needed. Define additional nozzle, water-conditioning or remote-monitoring functions separately from the standard oscillating-tube setup so the quotation and acceptance plan remain comparable.
Separate the supplied base equipment, selected options and site responsibilities in one configuration schedule. Agree the performance conditions and records for the selected build rather than treating every adjustable setting or accessory as a universal standard feature.
DR-H502-2
Choose the method and the evidence for your purchase
IPX1 IPX2 Drip Water Test Chamber for Mobile Phones · H503-2 IPX5 IPX6 Water Jet Test Chamber
Compare rain and water-ingress equipment
Choose equipment from the required exposure. Dripping, oscillating-tube spraying, water jets, immersion and close-range hot-water jets are different setups. Define additional apparatus explicitly if one purchase must cover more than one method.
For procurement and acceptance planning: environmental chamber buying guide · FAT checklist and blank acceptance record.
Prepare the IPX3 IPX4 Rain Spray Test Chamber DR-H502-2 enquiry
Send the following information together so the proposed equipment, accessories and acceptance scope match your test:
- Enclosure envelope
- IPX3 or IPX4 method
- tube coverage
- turntable load
- cable entry
- exposure and inspection sequence
List the required fixtures, measurement channels, data records and installation services separately in the quotation. Agree the representative specimen arrangement and pass criteria before the equipment acceptance run.
Check My Specimen Against the R400 Setup
Start with these details so the proposal can address your specimen, selected functions and acceptance requirements:
- Required IPX3/IPX4 apparatus and standard edition
- Full mounted envelope, mass and cable routing
- Tube radius, exposure positions and movement requirements
- Powered state and water-delivery verification
- Water, drain, electrical supply and required records
Check My Specimen Against the R400 Setup
Review DERUI company information and qualification documents. Ask which configuration records and documents apply to the proposed build.















