Floor 1, No. 3, Shugang Avenue, Hongmei Town, Dongguan City, Guangdong Province, China

20 years environmental test chamber manufacturer3000+ customer deliverables worldwide           Email: shirley@deruitest.com
WhatsApp Consultation:+86 135 2795 8363

IPX3 IPX4 Rain Spray Test Chamber DR-H502-2

Product overview

The DERUI DR-H502-2 uses an oscillating tube and rotating specimen support for spraying and splashing tests on electrical enclosures. Choose the tube and mounting arrangement around the complete assembly, not enclosure volume alone. A specimen-fit review checks moving clearance, exposed surfaces and the water-delivery verification required by your procedure.

Item Specification
Model / arrangement DR-H502-2 / oscillating-tube IPX3/IPX4 applications
Workspace (L × D × H) 900 × 900 × 900 mm
Tube radius R400 mm; R200 mm option listed
Table height / load 350–600 mm adjustment / 50 kg listed load
Equipment supply / power 220 V / 1.0 kW; destination connection to be specified
Safe and reliable
well-made
Quality assurance
customization
MessageWhatsApp
Hotline number 15580327593
  • Product Details
  • Technical parameters
  • FAQ
  • Contact us

DR-H502-2

IPX3 IPX4 Rain Spray Test Chamber DR-H502-2

The DERUI DR-H502-2 uses an oscillating tube and rotating specimen support for spraying and splashing tests on electrical enclosures. Choose the tube and mounting arrangement around the complete assembly, not enclosure volume alone. A specimen-fit review checks moving clearance, exposed surfaces and the water-delivery verification required by your procedure.

DR-H502-2

Choose the Spray Geometry Your Enclosure Needs

Repeatable spraying and splashing—not immersion

DR-H502-2 is intended for enclosure testing where the specified exposure is delivered by an oscillating tube. Applications can include electrical housings, outdoor instruments, lighting assemblies and components with seals or connections that need controlled water exposure.

The moving tube distributes water around the specimen. Table rotation changes the presented surfaces, and height adjustment helps position the assembly relative to the spray geometry. These mechanisms make mounting and coverage part of the test setup—not an afterthought.

Use the chamber to compare sealing designs during development and to execute a documented qualification procedure. Record the fixture and operating state as well as the water exposure so later tests can reproduce the same arrangement.

Start with the complete assembly: a cable gland or mounting bracket can change both the part’s clearance and the path water takes toward its seals.

Oscillating-tube spray test cabinet with specimen table and controls
DR-H502-2 IPX3 / IPX4 rain spray test chamber.

DR-H502-2

IPX3 and IPX4: Coverage Is More Than an Angle

Setup question IPX3 spraying water IPX4 splashing water
Exposure intent Spraying water within the prescribed angular region around the vertical Splashing water from the directions specified by the method
Spray arrangement Use the prescribed active-hole region and oscillating-tube setup Use the prescribed broader spray arrangement and specimen support
Motion Verify the selected sweep, speed/cycle and exposure time Verify the required sweep, speed/cycle and surface coverage
Specimen positioning Place the assembly at the required reference position without obstructing the spray Check support, orientation and whether all required surfaces receive exposure
Evaluation Apply the product-specific harmful-ingress and functional/electrical criteria Apply the product-specific harmful-ingress and functional/electrical criteria

IEC 60529 provides the enclosure IP-code framework. Its IPX3/IPX4 procedures distinguish the oscillating-tube arrangement from the spray-nozzle arrangement. Identify which apparatus your product standard or customer procedure requires.

Do not confuse spray coverage with tube travel. DR-H502-2 lists 120° for IPX3 and 350° for IPX4 as tube-angle settings. These are machine configuration values, not a substitute for checking the motion and water distribution required by the adopted standard. Define the angle convention and acceptance check in the purchase specification.

DR-H502-2

DR-H502-2 Technical Parameters

Parameter Configuration
Model DR-H502-2
Test arrangement Oscillating-tube spraying and splashing water for IPX3 / IPX4 applications
Internal workspace (L × D × H) 900 × 900 × 900 mm; nominal geometric volume 729 L
External dimensions (L × D × H) 1400 × 1120 × 1750 mm
Oscillating-tube radius R400 mm; R200 mm listed as an option
Listed tube travel 120° for IPX3; 350° for IPX4; custom settings available. Verify the travel definition and method-specific motion during acceptance
Specimen-table motion Continuous forward rotation or alternating forward/reverse rotation
Specimen-table height adjustment 350–600 mm
Listed table load 50 kg; include specimen and fixture in the loading review
Equipment supply / rated power 220 V / 1.0 kW; destination frequency and connection defined in the proposal
Listed specimen supply 220 V; electrical interface, allowable load and wet-test protection defined by the selected arrangement
Protection functions Electrical leakage protection, water-shortage protection and short-circuit protection
Listed warranty 1 year; coverage and service terms specified in the quotation

The 729 L figure is calculated from cabinet dimensions; usable specimen capacity is governed by the tube, support and clearances. For a method-matched configuration, agree spray-hole geometry, water delivery, motion cycle, table dimensions and timing requirements alongside these mechanical specifications.

DR-H502-2

Will Your Assembly Fit the R400 Spray Area?

Check the swept envelope

Include the fixture, cable bends, protruding connectors and the changing envelope during rotation. Keep the assembly clear of the moving tube across the complete programmed travel.

Align height and support

The listed table height is adjustable from 350 to 600 mm. Check the relevant specimen reference point against the tube center and use a support that follows the method without masking surfaces.

Account for fixture mass

Use the listed 50 kg table load for preliminary selection, then review combined specimen/fixture mass, mounting stability and off-center loading. Do not size the setup by specimen mass alone.

Fixture choices can change the result

A plate placed beneath the enclosure may shield the underside; a clamp may compress a gasket; an incorrectly routed cable may divert water away from a gland. The fixture should hold the specimen in the required orientation without improving its sealing performance or hiding an exposure surface.

If you want to test multiple units, submit a layout showing spacing, mounting and water access to each specimen. More parts fitting in the cabinet does not establish that each receives an acceptable exposure.

Ask for a specimen-fit review. Send a drawing or photograph with overall dimensions, mass, mounting orientation and cable route. Identify the seals or interfaces of interest so the proposal can address clearance and exposure—not only cabinet volume.

DR-H502-2

Water Delivery, Test Records and Factory Acceptance

Verify flow for the installed tube

Required total flow depends on the selected tube and active spray holes. Check the applicable method’s per-hole delivery and open-hole arrangement; total flow follows Q = N × q when N active holes each deliver q. Do not apply the same total flow to R200 and R400 without checking their geometry.

Before testing, inspect for blocked holes, confirm the flow indication and check tube motion and specimen position. A pressure reading alone cannot establish the correct water delivery. Follow the supplied water-quality and cleaning instructions to limit deposits and maintain the spray pattern.

Acceptance item What to agree and demonstrate
Tube and water distribution Installed radius, hole geometry and active-hole arrangement for each required exposure
Flow measurement Specified operating flow, instrument calibration and verification procedure
Motion and timing Angle definition, travel, cycle/speed and exposure timing at the agreed settings
Specimen setup Table diameter, usable height, loaded rotation, fixture clearance and required coverage
Electrical operation Equipment supply, sample supply/load, cable routing and protection checks
Records and documents Setup drawing, measured settings, test record format, manual, maintenance and agreed acceptance report

Separate exposure records from product approval

Record specimen ID, sealing configuration, mounting, test settings, water conditions, exposure duration and any interruptions. After testing, use the prescribed handling and inspection sequence; assess harmful ingress, operation and any electrical checks against the governing document.

Automatic logging, USB export, remote monitoring and water-temperature control should be specified separately if your laboratory needs them. For site planning, provide voltage/frequency, water supply, drain location, access route and service clearance. Maintain spray holes and filters where fitted, and never bypass wet-test electrical protection.

DR-H502-2

Compare the complete test setup before ordering

For purchasing

Compare the installed tube, active spray holes, total water delivery and full loaded motion. Include specimen-table dimensions and the rotating fixture, not just the 900mm cabinet. Ask each supplier to demonstrate the selected angle convention and motion recipe against your adopted procedure. The listed 120° and 350° settings require a method-matched acceptance check rather than an automatic compliance claim.

For the application engineer

For a housing with an offset cable gland, show the cable route and its swept envelope while the table turns. Check that clamps neither shelter the gland nor collide with the tube. For a batch, show every enclosure and support within the useful spray area; the number that fits on a shelf is not the number receiving an acceptable exposure.

Evidence to include

Agree measurements for the installed tube, hole arrangement, flow, motion and timing. Demonstrate the required sequence with the representative fixture and record any shielding or collision issue. Pressure indication alone is not a substitute for verifying the water delivered by the active holes.

DR-H502-2

Options to specify in the configuration proposal

Request interchangeable fixtures, a reviewed alternative tube, specimen monitoring and digital test records where needed. Define additional nozzle, water-conditioning or remote-monitoring functions separately from the standard oscillating-tube setup so the quotation and acceptance plan remain comparable.

Separate the supplied base equipment, selected options and site responsibilities in one configuration schedule. Agree the performance conditions and records for the selected build rather than treating every adjustable setting or accessory as a universal standard feature.

DR-H502-2

Choose the method and the evidence for your purchase

IPX1 IPX2 Drip Water Test Chamber for Mobile Phones · H503-2 IPX5 IPX6 Water Jet Test Chamber

Compare rain and water-ingress equipment

Choose equipment from the required exposure. Dripping, oscillating-tube spraying, water jets, immersion and close-range hot-water jets are different setups. Define additional apparatus explicitly if one purchase must cover more than one method.

For procurement and acceptance planning: environmental chamber buying guide · FAT checklist and blank acceptance record.

Prepare the IPX3 IPX4 Rain Spray Test Chamber DR-H502-2 enquiry

Send the following information together so the proposed equipment, accessories and acceptance scope match your test:

  • Enclosure envelope
  • IPX3 or IPX4 method
  • tube coverage
  • turntable load
  • cable entry
  • exposure and inspection sequence

List the required fixtures, measurement channels, data records and installation services separately in the quotation. Agree the representative specimen arrangement and pass criteria before the equipment acceptance run.

Check My Specimen Against the R400 Setup

Start with these details so the proposal can address your specimen, selected functions and acceptance requirements:

  • Required IPX3/IPX4 apparatus and standard edition
  • Full mounted envelope, mass and cable routing
  • Tube radius, exposure positions and movement requirements
  • Powered state and water-delivery verification
  • Water, drain, electrical supply and required records

Check My Specimen Against the R400 Setup

Review DERUI company information and qualification documents. Ask which configuration records and documents apply to the proposed build.

FacebookLinkedInXRedditWhatsApp

Technical parameters

Parameter Configuration
Model DR-H502-2
Test arrangement Oscillating-tube spraying and splashing water for IPX3 / IPX4 applications
Internal workspace (L × D × H) 900 × 900 × 900 mm; nominal geometric volume 729 L
External dimensions (L × D × H) 1400 × 1120 × 1750 mm
Oscillating-tube radius R400 mm; R200 mm listed as an option
Listed tube travel 120° for IPX3; 350° for IPX4; custom settings available. Verify the travel definition and method-specific motion during acceptance
Specimen-table motion Continuous forward rotation or alternating forward/reverse rotation
Specimen-table height adjustment 350–600 mm
Listed table load 50 kg; include specimen and fixture in the loading review
Equipment supply / rated power 220 V / 1.0 kW; destination frequency and connection defined in the proposal
Listed specimen supply 220 V; electrical interface, allowable load and wet-test protection defined by the selected arrangement
Protection functions Electrical leakage protection, water-shortage protection and short-circuit protection
Listed warranty 1 year; coverage and service terms specified in the quotation

Questions before ordering

Does DR-H502-2 perform both IPX3 and IPX4 tests?

It is offered for oscillating-tube IPX3/IPX4 exposure. Each test requires its own active spray-hole arrangement, flow, motion, duration and specimen support. Match those settings to the adopted procedure rather than changing only an angle on the controller.

Does a 900 mm chamber accept a 900 mm specimen?

No. The specimen and fixture must fit within the useful spray area and clear the R400 tube throughout its motion. Connectors, cables, table height and the specimen’s swept envelope also matter. Send the full assembly dimensions and mounting drawing for a fit review.

How is the R200 option different from R400?

R200 is a smaller-radius tube option, not an automatic capacity upgrade. The required radius depends on specimen size and the accepted test arrangement. A different tube also changes the active-hole configuration and corresponding total flow requirement.

Can specimens be energized during testing?

A 220 V specimen supply is listed. The proposal must define the outlet/interface, permitted electrical load, cable routing, leakage protection and shutdown procedure for the actual specimen. Other voltages or DC operation require a separately specified arrangement.

Is water pressure alone enough to verify spray severity?

No. Tube geometry, open spray holes, water delivery, motion, positioning and timing together define the exposure. Pressure is useful for operating checks but does not replace verification of the required flow and spray arrangement.

Will an IPX4 result demonstrate IPX6 or IPX8 resistance?

No. Water jets and immersion are different test arrangements. Select an IPX5/IPX6 jet chamber or an IPX7/IPX8 immersion system when those exposures are required. Evaluate the specimen according to its governing acceptance criteria; the machine itself does not issue a product certificate.

What should be included when comparing quotations?

Compare the installed tube, active spray holes, total water delivery and full loaded motion. Include specimen-table dimensions and the rotating fixture, not just the 900mm cabinet. Ask each supplier to demonstrate the selected angle convention and motion recipe against your adopted procedure. The listed 120° and 350° settings require a method-matched acceptance check rather than an automatic compliance claim.

Which project options should I specify?

Request interchangeable fixtures, a reviewed alternative tube, specimen monitoring and digital test records where needed. Define additional nozzle, water-conditioning or remote-monitoring functions separately from the standard oscillating-tube setup so the quotation and acceptance plan remain comparable.

WhatsAppWhatsApp
WhatsAppWhatsApp
WhatsApp
MessageMessage

Request a Quote