DERUI air-cooled xenon system
Air-Cooled Xenon Lamp Aging Test Chamber
Air-Cooled Xenon Lamp Aging Test Chamber supports separate light/dark moisture conditions and specimen irradiation area. Build the quotation around the specimen, programme and required measurements.
DERUI air-cooled xenon system
Air-Cooled Xenon Lamp Aging Test Chamber
Build a repeatable weathering program around the exposure your material actually needs. DERUI combines filtered xenon light with separately specified light/dark temperature and humidity conditions, plus programmable water spray, for comparative accelerated-aging evaluations.
DERUI air-cooled xenon system
What This Xenon Aging System Helps You Evaluate
Expose specimens consistently
Filtered xenon radiation, surface temperature and moisture act together on coatings, polymers and other exposed materials. Use an agreed cycle to compare formulations, suppliers or protective treatments under the same laboratory conditions.
Relevant outcomes include color change, gloss loss, cracking, embrittlement and retention of selected physical properties. The chamber produces the exposure; measurements such as tensile strength and color difference require the corresponding evaluation instruments.

DERUI air-cooled xenon system
Specify the Light, Dark and Spray Cycle
| Cycle element | What to define | Why it matters |
|---|---|---|
| Light exposure | Filter type, irradiance band/setpoint and exposure duration | A different spectrum or irradiance can change the degradation mechanism. Lux is not a substitute for method-specified irradiance. |
| Light-cycle climate | Chamber-air temperature, BPT or BST, and RH | The illuminated RH range is 20–60%; surface and air temperature are separate channels. |
| Dark-cycle climate | Duration, temperature and RH | Dark RH is specified separately at 30–98%; verify the selected combined operating points. |
| Water spray | Specimen spray timing, water quality, supply pressure and drainage | Wet/dry exposure must match the procedure; a rain interval alone does not define the full cycle. |
| Cycle repetition | Sequence, total exposure, inspection intervals and interruption rules | Consistent conditioning and restart rules improve comparability between runs. |
spray examples include 18 minutes spray / 102 minutes no spray and 12 minutes spray / 48 minutes no spray. These are programming examples, not interchangeable prescriptions for every ASTM or ISO method.
DERUI air-cooled xenon system
Configuration and Operating Ranges
Use these configuration values to shortlist the system. The selected equipment specification should identify the actual lamp rating, exposure area and achievable cycle conditions.
| Parameter | specification | Selection guidance |
|---|---|---|
| Chamber-air temperature | Ambient +10°C to 70°C | The lower limit depends on laboratory ambient temperature. This is not a sub-zero chamber range. |
| Black-panel temperature: light cycle | 40–100°C | Surface-temperature channel; not the chamber-air temperature. |
| Black-panel temperature: dark cycle | 25–85°C | Select achievable combinations with the air-temperature and humidity program. |
| Relative humidity: light cycle | 20–60% RH | Not a claim of 98% RH during illuminated exposure. |
| Relative humidity: dark cycle | 30–98% RH | The usable temperature/RH envelope must match the selected cycle. |
| Temperature uniformity | ±3°C | Specify the measuring channel, mapping points, load and operating condition for acceptance. |
| Lamp arrangement | Three air-cooled xenon lamps | Confirm the lamp arrangement in the selected configuration. |
| lamp-power options | 1.8 kW or 2.5 kW | Confirm the installed lamp rating. Lamp power is not total machine power. |
| irradiation-area options | 2,200–6,400 cm² | Usable specimen area depends on the selected chamber, holders and exposure geometry. |
| Spray duration setting | 0–9,999 minutes | Programmed duration; not a prescribed standard exposure. |
| Spray interval setting | 1–240 minutes | Select timing according to the adopted procedure. |
| Spray-water pressure | 0.12–0.15 MPa | Include water quality, supply and drainage in the installation plan. |
| Interior and specimen holders | SUS304 stainless-steel interior; stainless-steel holders | Confirm holder dimensions and fixture layout for your specimens. |
| Control | Programmable touchscreen; irradiance compensation | Agree the control wavelength, setpoint, calibration and recording requirements. |
DERUI air-cooled xenon system
Match the Apparatus to Your Test Procedure
| Reference | Primary purpose | Configuration to agree |
|---|---|---|
| ASTM G155 | Operation of xenon-arc light apparatus for exposure of materials | Adopted cycle, optical filters, irradiance control, temperature and moisture sequence. |
| ISO 4892-2 | Xenon-arc exposure methods for plastics | Required method and edition, specimen mounting, exposure conditions and evaluation plan. |
| ISO 16474-2 | Xenon-arc exposure methods for paints and varnishes | Coated-panel preparation, selected exposure method, reference specimens and endpoints. |
A standard name is not proof that every configuration supports every cycle. Include the exact procedure in the technical agreement, particularly the irradiance wavelength, sensor type and illuminated humidity requirement.
DERUI air-cooled xenon system
Plan Specimen Capacity, Utilities and Acceptance
Specimen layout
Send specimen dimensions, thickness, count and fixture drawings. Preserve the agreed exposure geometry and avoid mutual shading. The irradiation area is not automatically the usable loaded test area.
Laboratory installation
site guidance specifies ambient 5–28°C, a 24-hour average no higher than 28°C, RH no higher than 85%, and approximately 80 cm clearance around the machine. Confirm the final drawing, heat rejection, power, water and drainage before installation.
Acceptance evidence
Agree calibration for irradiance and temperature/RH channels, loaded specimen layout, cycle execution checks, spray verification, alarms and the recording/report format. Define tolerances and sensor locations before acceptance testing.
The control system includes programmable touchscreen operation and irradiance compensation. Door-open lamp shutoff is listed as a safeguard. Agree the supplied interlocks and alarm checks in the final equipment specification; never bypass UV-exposure protections.
For small flat-specimen workloads, review the desktop xenon chamber. For a different cooling architecture, review the water-cooled xenon weathering chamber. For material evaluation planning, see the paint and plastic xenon testing page. Compare the wider range in the xenon chamber category.
DERUI air-cooled xenon system
Compare the complete supply scope
Compare the complete optical configuration: lamp, filters, specimen exposure area, irradiance control, temperature measurement and selected moisture functions. Identify any spare lamp separately from operating lamps. Cooling water, purified spray water, room ventilation and electrical load must match the proposed build.
DERUI air-cooled xenon system
Choose the configuration for your application
Separate light/dark moisture conditions and specimen irradiation area.
Select daylight or window-glass filtering, specimen racks, irradiance control, moisture stages and recording according to the exposure programme. For solar-heating projects, define the illuminated area and uniformity, specimen heat balance and powered operation separately from accelerated weathering.
DERUI air-cooled xenon system
Agree the factory acceptance record
Witness the installed optical filter and lamp identification, irradiance calibration at the defined wavelength or band, temperature channels and exposure sequence. Check specimen placement and the agreed uniformity map. A lamp power or illuminance reading is not a substitute for spectral irradiance, and a material-weathering configuration is not automatically a classified photovoltaic solar simulator.
DERUI air-cooled xenon system
Compare related equipment
Prepare the Air-Cooled Xenon Lamp Aging Test Chamber enquiry
Send the following information together so the proposed equipment, accessories and acceptance scope match your test:
- Exposed panel area
- filter class
- light and dark setpoints
- spray and humidity steps
- room ventilation and electrical supply
List the required fixtures, measurement channels, data records and installation services separately in the quotation. Agree the representative specimen arrangement and pass criteria before the equipment acceptance run.
Request a air-cooled xenon configuration proposal
Send these details so the quotation covers your actual application:
- Exposed panel area
- Filter class
- Light and dark setpoints
- Spray and humidity steps
- Room ventilation and electrical supply
















