حجرة اختبار حرق البطارية

ال DERUI DR-D207 battery burning test chamber uses a Bunsen burner, stainless specimen screen and removable octagonal mesh cover for specified cell or small-battery flame-exposure procedures. Confirm the exact test clause and fixture drawing before ordering.

الموديل DR-D207
التركيز في الاختبار Controlled flame exposure with mesh surround; method-specific observation
Burner / Screen 10 mm burner mouth; 0.43 mm stainless wire, 20 openings per inch
Octagonal Cover 610 mm across opposite faces × 305 mm high; removable panels
Inner Size 750 × 750 × 500 mm (W × D × H)
Before Quotation Confirm clause/edition, cell energy and format, gas/exhaust, safety scope and FAT
آمن وموثوق
مصنوع بشكل جيد
ضمان الجودة
تخصيص
رسالةواتساب
رقم الخط الساخن 15580327593
  • تفاصيل المنتج
  • المعلمات الفنية
  • الأسئلة الشائعة
  • اتصل بنا
CELL FLAME EXPOSURE & PROJECTILE-CONTAINMENT SETUP

DR-D207 Battery Burning Test Chamber

A purpose-built burner, specimen screen and removable octagonal mesh cover for controlled flame-exposure testing of approved lithium cells or small batteries. Specify the adopted method, specimen and laboratory safety plan before treating any fixture geometry as a compliance claim.

What this equipment is: a flame-exposure fixture. It is not a programmable temperature chamber, a full-size pack fire tester or a UN 38.3 transport-test system.
الموديلDR-D207
Burner10 mm internal tube opening; approximately 100 mm long
cURL Too many subrequests by single Worker invocation. To configure this limit, refer to https://developers.cloudflare.com/workers/wrangler/configuration/#limits0.43 mm stainless wire; 20 openings per inch
Internal enclosure750 × 750 × 500 mm (W × D × H)

What the Test Helps You Observe

During a specified flame-exposure procedure, a cell or approved small battery sits above the burner while the screen and octagonal surround define the setup and help retain ejected material. The laboratory observes the behavior required by its adopted method; pass/fail criteria must come from that method and the test plan, not from the chamber specification alone.

Controlled setup

Published burner, screen, opening and octagonal-cover dimensions provide a starting point for method review.

Visible evidence

Agree in advance how the operator will record exposure time, specimen behavior and any ejected material.

Safer procurement

Confirm fuel supply, exhaust, operator protection and post-test handling before ordering.

Published DR-D207 Configuration

عنصر Published value cURL Too many subrequests by single Worker invocation. To configure this limit, refer to https://developers.cloudflare.com/workers/wrangler/configuration/#limits
Control and timer Button control; flame-time setting shown as 0–99H99M Confirm actual timer input format, resolution and shutoff behavior.
Burner and opening Bunsen burner, 10 mm internal mouth, approximately 100 mm length; test opening 100 ± 2 mm Confirm burner type, flame setting and fixture drawing against the adopted standard edition.
Specimen screen 0.43 mm stainless woven wire; 20 openings per inch Confirm mounting height and replacement screen supplied.
Octagonal cover 610 mm across opposite faces; 305 mm high; eight removable panels; 0.25 mm aluminum wire, 16–18 openings per inch Request dimensional drawing and panel-retention details.
cURL Too many subrequests by single Worker invocation. To configure this limit, refer to https://developers.cloudflare.com/workers/wrangler/configuration/#limits 750 × 750 × 500 mm internal; 1000 × 950 × 1310 mm external, including door lock Confirm usable clearance around the fixture and doorway.
المرافق Methane or LPG supplied by customer; 220 V, 50/60 Hz, 200 W; approximately 160 kg Confirm gas regulation, ventilation, site power and nameplate.
Relief and exhaust 200 × 200 mm pressure-relief door; rear Ø100 mm exhaust connection; door-retention chain Confirm discharge direction, duct requirements and risk-assessed containment limits.
Dimension requiring a drawing: the old specification states “height of the flame from the screen: 38 mm.” That wording does not unambiguously define the burner-to-screen geometry; verify it on the approved setup drawing.

Method Fit: Review the Clause, Not a Logo List

The DR-D207 screen and octagonal-surround geometry warrants review against the UL 1642 Projectile Test, but published fixture similarity is not proof of conformance. In particular, the DR-D207 page specifies a Bunsen burner, while a published technical comparison describes a Meker-type burner for the UL 1642 fire-exposure setup. Verify the adopted edition, burner, all dimensions, specimen preparation, observation and acceptance rules clause by clause before any UL 1642 claim.

Do not treat this chamber as a UN 38.3 test method. The UN transport manual has defined T.1–T.8 tests; a generic “UN 38.3 T.10 burning test” claim should not appear on this page. Other standards such as UL 2054 or IEC 62133-2 may form part of a laboratory program but do not automatically validate this fixture.

Send the exact method

Standard edition, clause, cell format, sample state, exposure period and pass/fail rule.

Confirm the physical setup

Burner specification, mesh, opening, screen-to-burner position and octagonal-cover drawing.

Confirm evidence

Video/observation plan, timed exposure, calibration records and final report format.

DR-D207 Is Different from DERUI's Other Battery Fire Equipment

Equipment Primary function Do not assume
DR-D207 burning test chamber Burner, screen and octagonal mesh fixture for specified cell/small-battery flame exposure Large pack fire testing, programmable heating or certified blast containment
DR-D211 EV battery cell fire tester A different external-flame test setup; compare its published specimen and burner limits separately Interchangeability with the DR-D207 projectile-test geometry
Battery thermal abuse chamber Controlled high-temperature exposure in a heated enclosure An open-flame burner or octagonal-mesh projectile-test fixture

Site Safety and Acceptance Plan

Review the specimen

Provide chemistry, dimensions, maximum energy, charge state and credible venting behavior. Agree allowable specimen size and quantity.

Review the facility

Confirm methane/LPG supply, regulator, exhaust route, clearance, fire response and operator exclusion area with site EHS.

Witness the setup

Check burner, screen and cover dimensions, timer, gas shutoff behavior, exhaust, door and documentation during FAT.

Safety boundary: the published relief door and retention chain are construction features, not evidence of a rated explosion-proof or thermal-runaway containment system. Gas detection, suppression, data export and interlocks are not established as standard features by the current public specification.

Questions Buyers Ask

Is this a UL-certified chamber?

No certification evidence is established by the public product page. Review fixture suitability against the exact UL 1642 edition and obtain any required third-party evaluation separately.

Can it test EV modules or packs?

Do not infer that from the enclosure size. Obtain a specimen-size, energy and hazard assessment, then compare the appropriate larger-scale equipment.

Is the gas included?

The published specification says methane or LPG is customer-provided. Confirm regulator, hose, shutoff and local installation scope in the quotation.

Does it perform UN 38.3 transport testing?

No. This burner arrangement should not be presented as a UN 38.3 test. Choose the dedicated equipment for the required T.1–T.8 method.

What needs confirmation before publication or purchase?

The adopted standard edition, fixture drawing, timer format, safety controls, real-device photos, accessories and measurable FAT criteria.

Request a Method-Matched DR-D207 Proposal

Send your cell drawing and test clause. DERUI can respond with the fixture drawing, included scope, utility list and an acceptance checklist that engineering and purchasing can review together.

Discuss the DR-D207 setup

Include in the inquiry

  • Standard edition and exact clause
  • Cell/battery format and energy
  • Gas and exhaust conditions
  • Required observation and FAT records
فيسبوكلينكدإنXريديتواتساب

DR-D207 values below come from the current DERUI product record. Confirm the approved drawing and safety configuration before ordering.

عنصر Published configuration / verification point
Model / control DR-D207; button control
Burner Bunsen burner; 10 mm internal tube-mouth diameter; approximately 100 mm tube length
Flame-time setting Published as 0–99H99M; confirm valid timer input format, resolution and shutoff logic
Test opening 100 ± 2 mm diameter
Specimen screen Stainless woven wire, 0.43 mm diameter; 20 openings per inch
Burner/screen geometry Published “38 mm” wording is ambiguous; require approved dimensioned drawing
Octagonal mesh cover 610 mm across opposite faces; 305 mm high; eight removable faces; 0.25 mm aluminum wire, 16–18 openings per inch
Fuel Methane or LPG; customer-provided. Confirm regulator, supply and shutoff requirements
Inner / outer dimensions 750 × 750 × 500 mm / 1000 × 950 × 1310 mm (W × D × H; outer includes door lock)
Relief / exhaust / door 200 × 200 mm relief door; rear Ø100 mm exhaust connection; left-opening door with retention chain. No rated containment claim established
Supply / weight 220 V single phase, 50/60 Hz, 200 W; approximately 160 kg

1. ما هو حجرة اختبار حرق البطارية؟
حجرة اختبار حرق البطارية هي حاوية مخصصة ومقواة بشكل كبير مصممة لاحتواء ودراسة حرائق البطاريات بأمان. تتيح للباحثين التسبب عمدًا في اندفاع حراري في خلايا الليثيوم أيون، مع مراقبة انتشار اللهب، وملامح درجة الحرارة، وفعالية الاحتواء تحت ظروف مسيطرة ومتكررة.

2. ما هي معايير السلامة التي يمكن تلبيةها باستخدام حجرة اختبار حرق البطارية؟
تم بناء هذه الحجرات لمساعدة المصنعين على الامتثال لمعايير السلامة الدولية الحيوية. من بين المعايير الرئيسية UL 9540A لأنظمة تخزين الطاقة،, UN 38.3 لنقل بطاريات الليثيوم،, ECE R100 للمركبات الكهربائية، و GB/T 31485 لمتطلبات سلامة بطاريات السيارات الكهربائية في مصر.

3. ما هي الميزات الأساسية للسلامة التي يجب البحث عنها؟
اعتمد على الحجرات التي تحتوي على أنظمة إخماد الحرائق التلقائية (مثل الرشاشات أو الغاز الخامل)،, تصميم مقاوم للانفجار, مراقبة الغاز ودرجة الحرارة في الوقت الحقيقي, و مفاتيح إيقاف الطوارئ. تعتبر هذه الميزات حاسمة لحماية المختبر والأفراد أثناء حرق مراقب.

4. ما هو الحد الأقصى لدرجة الحرارة التي يمكن لهذه الغرف تحملها؟
تم تصميم غرف اختبار حرق البطاريات عالية الجودة لتحمل الحرارة الشديدة الناتجة عن حريق البطارية. ابحث عن الطرازات التي تتمتع بدرجة حرارة تشغيل قصوى لا تقل عن 1100°C (2000°F) لضمان قدرتها على احتواء أخطر حالات الهروب الحراري دون تلف.

5. هل يمكن لنفس الغرفة اختبار أحجام وأنواع بطاريات مختلفة؟
نعم، تتميز العديد من الغرف الحديثة بـ تصميم معياري. يتيح ذلك إعادة تكوينها داخليًا أو استخدام تجهيزات مختلفة لاختبار كل شيء من خلايا الإلكترونيات الاستهلاكية الصغيرة إلى وحدات بطاريات السيارات الكهربائية أو أنظمة تخزين الطاقة الكبيرة.

6. ما نوع البيانات التي تجمعها الغرفة أثناء الاختبار؟
تم تجهيز الغرف المتقدمة بأنظمة جمع بيانات مدمجة تلتقط بيانات درجة حرارة عالية السرعة من عدة مقاييس حرارة, تغيرات الضغط, ومع الوحدات الاختيارية،, تحليل تركيب الغاز للغازات الناتجة أثناء الاحتراق.

7. كيف يتم تنظيف وصيانة الغرفة بعد اختبار الحريق؟
يتضمن تنظيف ما بعد الاختبار إزالة بقايا الحريق وفحص المكونات. الغرف التي تحتوي على داخلية من الفولاذ المقاوم للصدأ ناعمة, صواني الحطام القابلة للإزالة, و مجموعة رؤى قابلة للوصول تبسط بشكل كبير هذه العملية وتضمن جاهزيتها للاختبار التالي.

8. هل يمكن دمج الغرفة مع معدات اختبار أخرى؟
بالتأكيد. ابحث عن طرز مزودة بمنافذ اتصال قياسية (إيثرنت، CAN) تسمح لها بالمزامنة مع أجهزة غرف الاختبار البيئية للتكييف المسبق لدرجة الحرارة أو مع نظام إدارة البيانات المركزي في مختبرك لإعداد التقارير الموحدة.

11. ما هو الوقت النموذجي للتسليم والتركيب؟
نظرًا لطبيعتها المخصصة والشاقة، يمكن أن تتراوح المهل الزمنية لغرفة اختبار حرق البطاريات من 12 إلى 20 أسبوعًا. ويشمل ذلك التصنيع والتحقق من السلامة والتنسيق من أجل التركيب الاحترافي في الموقع وتدريب المشغلين.

10. هل تستخدم الغرفة لمراقبة الجودة أم فقط للبحث والتطوير؟
إنها تخدم كلا الوظيفتين. في البحث والتطوير, ، يستخدمها المهندسون للتحقق من صحة تصميمات الخلايا والمواد الجديدة. في مراقبة الجودة, مراقبة الجودة.

، يتم استخدامه لأخذ عينات الدفعات واختبارات الاعتماد لضمان استيفاء خلايا الإنتاج للوائح السلامة الإلزامية قبل الشحن.
11. ما نوع التهوية أو تجهيز المرافق المطلوب؟ تتطلب هذه الغرف نظام تهوية عادم قوي.

ومخصص لإزالة الأبخرة السامة والدخان بأمان بعد الاختبار. من الضروري التشاور مع الشركة المصنعة مبكرًا لفهم متطلبات الطاقة والمساحة والتهوية المحددة لمنشأتك.
12. ما هو الدعم والضمان المصاحب للغرفة؟ تركيب في الموقع, تدريب المشغل, التشخيص عن بعد, ، والوصول إلى شبكة عالمية لقطع الغيار للمكونات الهامة.

واتسابواتساب
واتسابواتساب
واتساب
رسالةرسالة

اطلب عرض سعر