Câmara de Teste de Temperatura e Humidade à Prova de Explosão de 3000L

A DERUI DR-H201-F3000 temperature and humidity test chamber is a non-standard 3,000 L system for large battery modules, packs and other high-load specimens. Climate performance, powered interfaces and the safety package are engineered from the specimen, test profile, credible hazards and installation site.

Modelo DR-H201-F3000
Nominal Volume 3,000 L; usable dimensions customized
Published Climate Range -70°C to +150°C; 10% to 98% RH
Project Basis Load, powered heat, climate profile, utilities and installation access
cURL Too many subrequests by single Worker invocation. To configure this limit, refer to https://developers.cloudflare.com/workers/wrangler/configuration/#limits Relief, detection, exhaust, suppression and interlocks require project confirmation
Before Production Approve drawings, performance conditions, safety matrix and FAT criteria
Seguro e fiável
bem construído
Garantia de qualidade
personalização
MensagemWhatsApp
Número de Linha Direta 15580327593
  • Detalhes do Produto
  • Perguntas Frequentes
  • Contacte-nos
Custom 3,000 L Battery Climate Chamber

DR-H201-F3000 Large Temperature and Humidity Chamber

Configure a 3,000 L controlled environment for battery modules, packs, energy-storage components and other large test loads. Temperature, humidity, specimen power, access ports and the safety response are engineered around the approved test plan and installation site.

Non-standard project: chamber dimensions, usable workspace, loaded performance, interfaces and safety functions are not fixed by volume alone. They must be frozen in the technical agreement before production.
ModeloDR-H201-F3000
Published volume3,000 L
cURL Too many subrequests by single Worker invocation. To configure this limit, refer to https://developers.cloudflare.com/workers/wrangler/configuration/#limits-70°C a +150°C
Published humidity rangeRH 10% a 98%

Designed Around Large Loads, Not an Empty Chamber

Battery modules and packs

Provide specimen envelope, mass, state of charge, maximum electrical energy and powered heat so airflow and refrigeration capacity can be sized correctly.

Multiple specimens or fixtures

Define rack layout, clearances, cable routing and loading method. A nominal 3,000 L volume does not equal 3,000 L of usable specimen space.

Integrated test systems

Specify cycler connections, voltage and temperature channels, gas sampling, video, exhaust and data interfaces before the chamber layout is released.

Published DR-H201-F3000 Configuration

Item Published value Project confirmation
Model / nominal volume DR-H201-F3000 / 3,000 L Confirm usable internal dimensions and door opening on the approved drawing.
Faixa de temperatura -70°C a +150°C Confirm required option, load, setpoint span and stabilization criteria.
Uniformidade de temperatura ±2°C Confirm whether this means deviation or uniformity, sensor map and loaded/no-load condition.
Faixa de humidade RH 10% a 98% Request the combined temperature-humidity operating envelope; full RH range is not available at every temperature.
Heating / cooling rate Published as up to 5°C/min / 3°C/min Define average or linear rate, start/end points, specimen load and cooling-water condition.
Fonte de alimentação 220 V or 380 V, 50/60 Hz; customized Freeze voltage, phase, frequency, connected load and protective device.
Internal / external material Stainless-steel interior / reinforced coated-steel exterior Confirm grade, thickness, floor loading and corrosion requirements.
External dimensions Customized for installation Confirm transport route, doorway, service clearance and maintenance access.

Freeze the Non-Standard Design Inputs First

Define the climate duty

List temperature and RH setpoints, transitions, dwell times, cycles, tolerances and recovery requirements.

Define the real load

Provide dimensions, mass, quantity, fixtures, packaging, powered heat, cable loss and expected failure modes.

Define the facility interface

Confirm electrical supply, cooling water, drain, exhaust route, fire-system interface, room ambient and installation access.

Safety Functions Must Follow the Hazard Analysis

Function Information required before selection Acceptance evidence
cURL Too many subrequests by single Worker invocation. To configure this limit, refer to https://developers.cloudflare.com/workers/wrangler/configuration/#limits Credible pressure event, relief direction, safe discharge area and facility clearance Approved calculation/drawing and functional inspection
Gas, smoke or flame detection Target hazard, sensor type/range, alarm thresholds and calibration Calibration record and cause-and-effect test
Forced exhaust Required airflow, duct material, discharge point, make-up air and purge time Measured airflow and interlock demonstration
Fire suppression Agent, design concentration, activation logic, drainage and local-code approval Approved system documents and functional test
cURL Too many subrequests by single Worker invocation. To configure this limit, refer to https://developers.cloudflare.com/workers/wrangler/configuration/#limits Specimen power, chamber shutdown sequence, emergency stop and fail-safe state Interlock matrix and witnessed shutdown test
Reinforced enclosure Defined event, structural design basis, door/window/port construction Drawing review and agreed project-specific verification
Safety wording: “explosion-proof” is not a universal guarantee of battery thermal-runaway containment. The quotation must identify the exact functions supplied and the event they are designed to address.

DR-H201-F3000 Reference Image

Custom 3000L battery temperature and humidity test chamber with battery pack fixture

Reference image from the current product page. Final internal layout, access ports, external services and safety package follow the approved project drawing.

Standards and Certification Need Separate Decisions

Reference How it may relate What must be verified
IEC 60068-2-1 / IEC 60068-2-2 Cold and dry-heat environmental tests Applicable procedure, severity, load and chamber performance.
IEC 60068-2-78 Cab: steady-state damp heat Temperature-RH point, duration, specimen state and acceptance criteria.
UN 38.3 Lithium battery transport-test programme Only the applicable procedure should be mapped; the chamber does not perform every T.1–T.8 test.
UL 1642 / UL 2580 or customer specification Method-specific cell or EV battery safety evaluation Exact edition, clause, fixtures, monitoring and separate abuse equipment.
ATEX / IECEx / IEC 60079 Equipment for potentially explosive atmospheres Do not claim certification without the exact certificate, model/configuration scope, equipment marking and issuing body.

Recommended FAT for a Custom 3,000 L Chamber

Climate performance

Verify agreed temperature/RH points, empty and representative-load mapping, ramp definition, recovery and door-opening behaviour.

cURL Too many subrequests by single Worker invocation. To configure this limit, refer to https://developers.cloudflare.com/workers/wrangler/configuration/#limits

Check programme execution, alarms, access permissions, sensor records, export format, communication and power-failure recovery.

Safety cause and effect

Simulate every detector and fault input to verify alarms, isolation, exhaust, suppression command and safe-state sequence.

FAT limitation: a functional interlock test is not the same as proving containment of an uncontrolled battery event. Any destructive or certification test requires a separately agreed protocol and facility.

Information Required for an Accurate Quotation

Topic Send DERUI
Specimen Dimensions, mass, quantity, chemistry, capacity, SOC, electrical energy and packaging.
Test profile Temperature/RH programme, rates, dwell, cycles, tolerance and specimen operating mode.
Powered load Maximum watts, charge/discharge current, duty cycle and cable/feedthrough requirements.
Failure scenario Credible gas, smoke, flame, pressure, electrolyte, fragment and heat-release conditions.
Facility Power, cooling water, drainage, exhaust route, fire system, ambient and installation access.
cURL Too many subrequests by single Worker invocation. To configure this limit, refer to https://developers.cloudflare.com/workers/wrangler/configuration/#limits Mapping points, loaded performance, alarms, interlocks, data format and documentation.

Perguntas Frequentes

Is DR-H201-F3000 a standard catalogue chamber?

No. It is a non-standard 3,000 L project. Dimensions, refrigeration, humidification, ports, utilities and safety controls must be confirmed against the test load and site.

Is 10% to 98% RH available throughout -70°C to +150°C?

No such full-envelope capability is established. Request the combined temperature-RH chart and define the required operating points.

Is the published ramp rate guaranteed with a battery pack installed?

Not without a defined load. Rate type, span, payload mass, powered heat and facility conditions must be written into the technical agreement.

Does “explosion-proof” guarantee containment of thermal runaway?

No. A specific hazard scenario, engineered protection package, acceptance protocol and supporting evidence are required.

Is this model confirmed ATEX or IECEx certified?

The current page makes that claim but provides no certificate number or scope. Treat certification as unverified until DERUI supplies the certificate covering the quoted model and configuration.

What should be approved before production?

Approve the chamber drawing, usable workspace, climate duty, load, interfaces, utilities, safety matrix, FAT criteria, documentation and delivery scope.

Start with the Load, Hazard and Site—Then Freeze the Chamber

Send the 3,000 L project profile, specimen data, powered heat, failure scenarios and facility utilities. DERUI can return a configuration matrix and measurable FAT scope for technical review.

Request a 3,000 L Project Review

FacebookLinkedInXRedditWhatsApp

1. P: Para que serve uma câmara de teste de temperatura e humidade à prova de explosões de 3000L?
A: Esta câmara é principalmente projetada para testes de fiabilidade seguros de produtos perigosos em uma atmosfera controlada e explosiva. A sua grande capacidade de 3000L torna-a ideal para testar pacotes de baterias de íons de lítio em grande quantidade, sistemas de armazenamento de energia (ESS) e materiais químicos perigosos sob condições extremas de temperatura (-70°C a +150°C) e humidade (10-98% HR), garantindo que não inflamem nem explodam durante o stress térmico, ciclos de carga/descarga ou simulação ambiental.

2. P: Como é que uma câmara de teste à prova de explosões garante segurança durante o teste de baterias?
A: A nossa câmara à prova de explosões de 3000L incorpora um design de segurança em múltiplas camadas. Apresenta ventilações de alívio de pressão, componentes elétricos à prova de explosões e construção reforçada para conter e ventilar com segurança qualquer evento interno de fuga térmica. Integrados detectores de fumaça, sensores de chama e um sistema automático de supressão de incêndios acionam imediatamente o desligamento e a supressão para proteger tanto a amostra quanto o ambiente do laboratório.

3. P: Quais certificações uma câmara de clima à prova de explosões confiável precisa ter?
A: Para aceitação no mercado global, uma câmara profissional deve possuir certificações ATEX e IECEx para uso em atmosferas explosivas. Além disso, o cumprimento de normas de segurança de baterias como UN 38.3, UL 1642 e UL 2580 é fundamental. A nossa câmara de 3000L atende a essas certificações, fornecendo prova legalmente reconhecida de segurança para validação de produtos e conformidade de transporte.

4. P: Por que escolher uma câmara de grande capacidade de 3000L para testes à prova de explosões?
A: A Espaço de trabalho de 3000L é essencial para testar módulos de bateria de tamanho completo, packs ou múltiplos componentes simultaneamente, replicando condições reais de montagem. Isso elimina a necessidade de testes em escala reduzida, fornecendo dados de segurança mais precisos e confiáveis para veículos elétricos (VE), aeroespacial e aplicações industriais de armazenamento de energia.

5. P: Qual é a faixa de temperatura e humidade desta câmara, e por que é importante?
A: Esta câmara oferece uma faixa ampla de -70°C a +150°C e 10% a 98% de humidade relativa. Esta amplitude é fundamental para simular extremos climáticos globais— desde frio ártico até calor desértico e alta humidade — e para realizar testes rigorosos de ciclo térmico e stress em baterias e eletrónica para identificar pontos de falha e garantir durabilidade a longo prazo.

6. P: Esta câmara pode ser personalizada para necessidades específicas de teste?
A: Com certeza. Oferecemos opções extensas de personalização incluindo portos adicionais para sensores, integração com ciclistas de baterias ou mesas de vibração, prateleiras especializadas e perfis de temperatura/humidade personalizados. Isto garante que a câmara se encaixe perfeitamente na sua configuração específica de bancada de testes para protocolos complexos de testes de fiabilidade multifatoriais.

7. P: Como é controlada e monitorizada a câmara?
A: Dispõe de um controlador lógico programável (PLC) de fácil utilização com interface de ecrã tátil. Os operadores podem definir perfis complexos, enquanto monitorização remota em tempo real e registo de dados permitem operação sem supervisão e registo preciso das condições de teste, o que é vital para rastreabilidade e relatórios.

8. P: Quais indústrias mais necessitam deste tipo de câmara de teste ambiental à prova de explosões?
A: Indústrias principais incluem:

  • Veículos de Nova Energia (EV): Para validação de segurança de packs de baterias e BMS.

  • Aeroespacial e Defesa: Para testes de baterias e componentes em ambientes severos e críticos de segurança.

  • Química e Farmacêutica: Para testes de estabilidade de substâncias voláteis ou reativas.

  • Eletrónica de Consumo: Para testes rigorosos de segurança de baterias de alta capacidade em dispositivos.

9. P: Quais são as principais diferenças entre uma câmara ambiental padrão e uma à prova de explosões?
A: As principais diferenças estão em engenharia de segurança. Uma câmara à prova de explosões possui estrutura reforçada, mecanismos de alívio de pressão, eletrónica intrinsecamente segura, ventiladores à prova de faísca e sistemas avançados de deteção de gases/extinção de incêndios. É construída para conter uma explosão interna e prevenir a ignição da atmosfera externa do laboratório, enquanto uma câmara padrão não é.

10. P: O que devo procurar num fabricante de grandes câmaras de teste à prova de explosões?
A: Escolha um fabricante com:

  • Experiência Comprovada: Experiência demonstrada na construção de equipamentos certificados à prova de explosões.

  • Materiais de Qualidade: Uso de interiores em aço inoxidável e componentes robustos.

  • Suporte Abrangente: Oferece instalação, formação, calibração e serviço pós-venda local.

  • Conhecimento de Conformidade: Compreensão profunda de ATEX, IECEx e normas relevantes da indústria para garantir que os seus testes estejam em conformidade e sejam reconhecidos.