The DERUI DR-D203 battery altitude simulation test chamber provides controlled low-pressure exposure for lithium cell and battery testing. Its 125 L workspace and 5.0–101.3 kPa absolute-pressure range include the 11.6 kPa condition used in UN 38.3 T.1. This is a low-pressure-only chamber; the laboratory maintains and records the required ambient temperature separately.

МодельDR-D203 battery low-pressure chamber
Workspace500 × 500 × 500 mm (W × D × H), 125 L
Диапазон давления5.0–101.3 kPa absolute; UN 38.3 T.1 reference condition: 11.6 kPa or less
ControlE600 7-inch touchscreen; automatic pressure and hold-time settings; time-stamped pressure recording and export
ТемператураNo built-in heating or cooling; laboratory ambient condition is maintained and documented separately
ПрименениеBattery low-pressure exposure and UN 38.3 T.1 test planning; the complete test and report remain the laboratory’s responsibility

Безопасно и надежно
качественно изготовлено
Гарантия качества
кастомизация
СообщениеWhatsApp
Номер горячей линии 15580327593
  • Детали продукта
  • Технические параметры
  • Часто задаваемые вопросы
  • Свяжитесь с нами
DR-D203 · Battery Low-Pressure Chamber

125 L Battery Altitude Simulation Test Chamber

Specify a controlled absolute-pressure hold for battery transport testing without paying for an unneeded heating and refrigeration system. The revised DR-D203 has a 500 × 500 × 500 mm workspace, a 5.0–101.3 kPa absolute-pressure range and automatic vacuum control through a 7-inch touchscreen.

Workspace500 × 500 × 500 mm · 125 L
Absolute pressure5.0–101.3 kPa
Test focusLow-pressure exposure, including the 11.6 kPa T.1 condition

What this chamber is—and is not

DR-D203 is a low-pressure altitude simulation chamber for cells, batteries and other specimens whose test plan calls for reduced air pressure. It does not include active heating or refrigeration. For UN 38.3 T.1, the laboratory must maintain and document the required 20 ± 5 °C ambient condition independently; a pressure-capable chamber alone does not certify a test result.

Do not substitute this model for a temperature-altitude chamber. Simultaneous programmed hot/cold and low-pressure exposure requires a separately specified system. The observation window and door restraint are standard construction features, not a verified thermal-runaway containment rating.

DR-D203 technical specifications

Параметр 125 L configuration
Internal workspace 500 × 500 × 500 mm (W × D × H), 125 L geometric volume; two levels with two mesh shelves
cURL Too many subrequests by single Worker invocation. To configure this limit, refer to https://developers.cloudflare.com/workers/wrangler/configuration/#limits Approximately 800 × 750 × 1480 mm (W × D × H); final unit governs
Диапазон давления 5.0–101.3 kPa absolute; minimum 5.0 kPa
Pressure control accuracy At ≤5 kPa: ±0.5 kPa; above 5–40 kPa: ±1 kPa; above 40–80 kPa: ±2 kPa. Confirm the agreed acceptance point and test conditions in the order.
Pressure release rate 15 ± 3 kPa/min
Control and sensing E600 Youyi 7-inch touchscreen; diffused-silicon pressure sensor; automatic upper/lower pressure limits, hold time, end-of-test venting and alarm settings
Data record Time-stamped pressure curve and export included in the confirmed standard configuration; request the file format, sampling interval and FAT demonstration in the order.
Chamber structure 4.0 mm SUS304 stainless-steel reinforced interior; 1.2 mm powder-coated cold-rolled steel exterior; rock-wool infill; high-temperature silicone door seal
Window and door 19 mm tempered-glass window, 250 × 300 mm; door restraint chain
Vacuum pump Daertuo VP750; 220 V / 0.25 kW motor; pump noise ≤75 dB as supplied specification
cURL Too many subrequests by single Worker invocation. To configure this limit, refer to https://developers.cloudflare.com/workers/wrangler/configuration/#limits Emergency stop, pressure-limit protection, pressure-reducing valve, automatic and manual venting, grounding and fast fuse; protection scope must be checked against the actual battery hazard assessment
Power and weight 220 V / 50 Hz, rated equipment power 2 kW; approximately 250 kg
Обслуживание Replaceable custom door seal; vacuum-pump oil replacement and routine pump service

Product quotation, drawings and factory acceptance results govern the delivered unit.

How to specify a UN 38.3 T.1 setup

Test profileIdentify the adopted Manual edition, absolute-pressure target at or below 11.6 kPa, required hold, test sequence and laboratory ambient-temperature record.
cURL Too many subrequests by single Worker invocation. To configure this limit, refer to https://developers.cloudflare.com/workers/wrangler/configuration/#limitsProvide chemistry, dimensions, mass, quantity, state of charge and condition. Shelf load limits and usable clearance must be approved before loading.
cURL Too many subrequests by single Worker invocation. To configure this limit, refer to https://developers.cloudflare.com/workers/wrangler/configuration/#limitsAgree on pressure/temperature instruments, calibration evidence, exported traces, observations and before/after measurements needed by the laboratory.

This equipment supports the low-pressure portion of a test program; it does not itself confer UN 38.3 approval. Confirm the applicable method and reporting criteria with your laboratory.

Standard safety equipment and operating limits

The confirmed standard configuration includes an emergency stop, pressure protection, pressure-reducing valve, automatic end-of-test release and manual venting. A 19 mm tempered-glass window and side-mounted door restraint are also fitted. The vacuum pump has an exhaust outlet. None of these statements establishes a rated capability to contain battery fire, explosion or thermal runaway.

The standard chamber has no confirmed powered-test feedthrough. Keep specimens unpowered during a standard T.1 storage exposure unless a separately engineered configuration and approved procedure explicitly require otherwise. Gas detection, fire interfaces and laboratory exhaust treatment must be evaluated for the specimen and facility; do not assume they are included.

Real DR-D203 equipment photograph

UN 38.3 T.1 lithium battery altitude simulation workflow from low pressure exposure to result evaluation

Common selection questions

Does the DR-D203 control temperature?

No. It is a low-pressure-only chamber without active heating or cooling. The laboratory must establish and document the ambient condition required by its test procedure.

Is a 5.0 kPa minimum enough for UN 38.3 T.1?

The specified pressure range includes the T.1 target of 11.6 kPa absolute or less. Whether a particular test is valid also depends on duration, ambient condition, samples, measurements and the adopted method edition.

Can the chamber generate a compliant test report by itself?

No. Its standard pressure-trace export supports evidence collection, while the laboratory remains responsible for calibration, sample identification, observations, electrical measurements and the complete report.

Configure the chamber for your test plan

Send the required standard edition, specimen details, pressure profile, lab ambient conditions and reporting requirements. DERUI can confirm the delivered configuration, shelf capacity, safety scope and factory acceptance points before quotation.

Request a DR-D203 configuration review

FacebookLinkedInXRedditWhatsApp
Модель DR-D203
Размер внутренней коробки 400*400*400мм (Ш*Г*В)
Размер оборудования 630*620*1130мм (Ш*Г*В)
Материал внутренней коробки Нержавеющая сталь SUS304, толщина 3,0мм
Материал внешней коробки Холоднокатаная стальная листовая поверхность с покрытием, толщина 1,5мм
Окно для просмотра Размер 250*250мм, двухслойное закаленное стекло, прозрачное окно с сеткой из нержавеющей стали
Мобильность оборудования На дне оборудования расположены четыре универсальных ролика, что позволяет свободно перемещать его
Диапазон давления 0,5-100 кПа, обычно используемое тестовое давление воздуха 11,6 кПа
Отображение уровня вакуума Цифровой дисплей SMC
Метод управления Управление кнопками
Таймер Цифровой дисплей LED, настройка в диапазоне 0-9999 часов
Колебания давления ≤5%
Дверь ящика Однодверный с левым открыванием и взрывозащитной цепью сбоку двери
Лоток для хранения 2
Вес оборудования 130 кг
Электропитание AC220V
Мощность 1 кВт

1.Что такое камера испытаний моделирования высоты батареи?
Это специализированная камера, предназначенная для тестирования батарей в различных моделируемых условиях высоты.

2.Как работает моделирование высоты?
Камера использует точные датчики и системы управления для воспроизведения различных условий высоты для точного тестирования.

3.Почему тестирование на высоте важно для батарей?
Тестирование на высоте важно для понимания производительности и надежности батарей в различных экологических условиях.

4.Какие типы батарей можно тестировать?
Камера универсальна и может тестировать различные типы батарей, включая литий-ионные, никель-металгидридные и другие.

5.Включено ли управление температурой?
Да, наша камера включает передовое тепловое управление для поддержания оптимальных условий тестирования.

6.Могу ли я настроить профили тестирования?
Конечно! Камера позволяет создавать индивидуальные конфигурации тестов, адаптированные к конкретным типам батарей.

7, Какие функции безопасности включены?
Механизмы безопасности включают системы аварийного отключения, защиту от избыточного давления и комплексную вентиляцию.

8, Является ли установка сложной?
Установка проста, и наши эксперты предоставляют поддержку для обеспечения правильной настройки.