IPX9K Test Chamber | High-Pressure Hot-Water Testing

Test Application: High-pressure, high-temperature water-jet ingress testing for exposed enclosures and components

Test Designation: IPX9K; IEC IPX9 capability should be confirmed against the required standard edition and test method

نظام الرش: Four fixed angles at 0°, 30°, 60° and 90°

Published Water Conditions: 8–10 MPa pressure | 14–16 L/min flow | Water temperature up to 80°C

Turntable: Ø400 mm | 4–10 rpm | 50 kg rated load

Payment: L/C, D/P or T/T

Delivery: Confirmed after chamber configuration, specimen requirements and test method are approved

آمن وموثوق
مصنوع بشكل جيد
ضمان الجودة
تخصيص
رسالةواتساب
رقم الخط الساخن 15580327593
  • تفاصيل المنتج
  • المعلمة
  • الأسئلة الشائعة
  • اتصل بنا

High-Pressure Hot-Water Ingress Testing in a Controlled Chamber

The DERUI IPX9K test chamber evaluates whether an enclosure can resist close-range, high-pressure hot-water jets. Its four-angle spray arrangement, controlled water conditions and rotating specimen table support repeatable testing of automotive components, electrical enclosures, outdoor equipment and other products exposed to severe wash-down conditions.

0° / 30° / 60° / 90° jets8–10 MPa published pressureUp to 80°C waterProgrammable turntable

IPX9 and IPX9K Are Not Labels to Use Interchangeably

ISO 20653 addresses enclosure protection for electrical equipment in road vehicles and uses the IPX9K designation. IEC 60529 defines enclosure protection for electrical equipment and includes an IPX9 high-pressure, high-temperature water-jet test in current editions. The exact fixture, exposure conditions and acceptance criteria must come from the standard edition, customer specification or vehicle-manufacturer procedure named in the test plan.

Before production, DERUI reviews the required method rather than assuming that an “IPX9K chamber” automatically satisfies every IPX9 or OEM procedure.

How the IPX9K Test Chamber Works

Four-Angle Water-Jet Exposure

Nozzles positioned at 0°, 30°, 60° and 90° direct hot, pressurized water at the specimen. The required nozzle geometry and distance are set during commissioning.

Controlled Pressure, Flow and Temperature

A pump, heater and control system regulate the three variables that most strongly affect exposure severity. These channels should be verified before formal testing.

Rotating Specimen Platform

The specimen turns at a controlled speed so that exposed surfaces receive a repeatable spray sequence. Fixtures must hold the sample securely without masking critical seals.

Enclosed, Safer Operation

The stainless-steel test space contains splash and hot water. Door interlock, over-temperature protection, low-water protection and emergency stop options reduce operator risk.

Published Chamber Configuration

The following values consolidate the configuration currently published for this DERUI page. The approved quotation and technical agreement remain controlling because chamber dimensions, fixture design, utilities and safety options can change with the specimen.

Internal workspace 800 × 800 × 800 mm
Approximate external dimensions cURL Too many subrequests by single Worker invocation. To configure this limit, refer to https://developers.cloudflare.com/workers/wrangler/configuration/#limits
cURL Too many subrequests by single Worker invocation. To configure this limit, refer to https://developers.cloudflare.com/workers/wrangler/configuration/#limits cURL Too many subrequests by single Worker invocation. To configure this limit, refer to https://developers.cloudflare.com/workers/wrangler/configuration/#limits
cURL Too many subrequests by single Worker invocation. To configure this limit, refer to https://developers.cloudflare.com/workers/wrangler/configuration/#limits cURL Too many subrequests by single Worker invocation. To configure this limit, refer to https://developers.cloudflare.com/workers/wrangler/configuration/#limits
cURL Too many subrequests by single Worker invocation. To configure this limit, refer to https://developers.cloudflare.com/workers/wrangler/configuration/#limits cURL Too many subrequests by single Worker invocation. To configure this limit, refer to https://developers.cloudflare.com/workers/wrangler/configuration/#limits
نطاق الضغط cURL Too many subrequests by single Worker invocation. To configure this limit, refer to https://developers.cloudflare.com/workers/wrangler/configuration/#limits
cURL Too many subrequests by single Worker invocation. To configure this limit, refer to https://developers.cloudflare.com/workers/wrangler/configuration/#limits cURL Too many subrequests by single Worker invocation. To configure this limit, refer to https://developers.cloudflare.com/workers/wrangler/configuration/#limits
cURL Too many subrequests by single Worker invocation. To configure this limit, refer to https://developers.cloudflare.com/workers/wrangler/configuration/#limits cURL Too many subrequests by single Worker invocation. To configure this limit, refer to https://developers.cloudflare.com/workers/wrangler/configuration/#limits
cURL Too many subrequests by single Worker invocation. To configure this limit, refer to https://developers.cloudflare.com/workers/wrangler/configuration/#limits cURL Too many subrequests by single Worker invocation. To configure this limit, refer to https://developers.cloudflare.com/workers/wrangler/configuration/#limits
مصدر الطاقة cURL Too many subrequests by single Worker invocation. To configure this limit, refer to https://developers.cloudflare.com/workers/wrangler/configuration/#limits.
مياه الإمداد cURL Too many subrequests by single Worker invocation. To configure this limit, refer to https://developers.cloudflare.com/workers/wrangler/configuration/#limits.
cURL Too many subrequests by single Worker invocation. To configure this limit, refer to https://developers.cloudflare.com/workers/wrangler/configuration/#limits cURL Too many subrequests by single Worker invocation. To configure this limit, refer to https://developers.cloudflare.com/workers/wrangler/configuration/#limits.

cURL Too many subrequests by single Worker invocation. To configure this limit, refer to https://developers.cloudflare.com/workers/wrangler/configuration/#limits

cURL Too many subrequests by single Worker invocation. To configure this limit, refer to https://developers.cloudflare.com/workers/wrangler/configuration/#limits cURL Too many subrequests by single Worker invocation. To configure this limit, refer to https://developers.cloudflare.com/workers/wrangler/configuration/#limits cURL Too many subrequests by single Worker invocation. To configure this limit, refer to https://developers.cloudflare.com/workers/wrangler/configuration/#limits Equipment implication
cURL Too many subrequests by single Worker invocation. To configure this limit, refer to https://developers.cloudflare.com/workers/wrangler/configuration/#limits Water jets cURL Too many subrequests by single Worker invocation. To configure this limit, refer to https://developers.cloudflare.com/workers/wrangler/configuration/#limits cURL Too many subrequests by single Worker invocation. To configure this limit, refer to https://developers.cloudflare.com/workers/wrangler/configuration/#limits
cURL Too many subrequests by single Worker invocation. To configure this limit, refer to https://developers.cloudflare.com/workers/wrangler/configuration/#limits cURL Too many subrequests by single Worker invocation. To configure this limit, refer to https://developers.cloudflare.com/workers/wrangler/configuration/#limits cURL Too many subrequests by single Worker invocation. To configure this limit, refer to https://developers.cloudflare.com/workers/wrangler/configuration/#limits cURL Too many subrequests by single Worker invocation. To configure this limit, refer to https://developers.cloudflare.com/workers/wrangler/configuration/#limits
cURL Too many subrequests by single Worker invocation. To configure this limit, refer to https://developers.cloudflare.com/workers/wrangler/configuration/#limits cURL Too many subrequests by single Worker invocation. To configure this limit, refer to https://developers.cloudflare.com/workers/wrangler/configuration/#limits cURL Too many subrequests by single Worker invocation. To configure this limit, refer to https://developers.cloudflare.com/workers/wrangler/configuration/#limits cURL Too many subrequests by single Worker invocation. To configure this limit, refer to https://developers.cloudflare.com/workers/wrangler/configuration/#limits

cURL Too many subrequests by single Worker invocation. To configure this limit, refer to https://developers.cloudflare.com/workers/wrangler/configuration/#limits.

cURL Too many subrequests by single Worker invocation. To configure this limit, refer to https://developers.cloudflare.com/workers/wrangler/configuration/#limits

cURL Too many subrequests by single Worker invocation. To configure this limit, refer to https://developers.cloudflare.com/workers/wrangler/configuration/#limits

cURL Too many subrequests by single Worker invocation. To configure this limit, refer to https://developers.cloudflare.com/workers/wrangler/configuration/#limits.

cURL Too many subrequests by single Worker invocation. To configure this limit, refer to https://developers.cloudflare.com/workers/wrangler/configuration/#limits

cURL Too many subrequests by single Worker invocation. To configure this limit, refer to https://developers.cloudflare.com/workers/wrangler/configuration/#limits.

cURL Too many subrequests by single Worker invocation. To configure this limit, refer to https://developers.cloudflare.com/workers/wrangler/configuration/#limits

cURL Too many subrequests by single Worker invocation. To configure this limit, refer to https://developers.cloudflare.com/workers/wrangler/configuration/#limits.

cURL Too many subrequests by single Worker invocation. To configure this limit, refer to https://developers.cloudflare.com/workers/wrangler/configuration/#limits

cURL Too many subrequests by single Worker invocation. To configure this limit, refer to https://developers.cloudflare.com/workers/wrangler/configuration/#limits.

cURL Too many subrequests by single Worker invocation. To configure this limit, refer to https://developers.cloudflare.com/workers/wrangler/configuration/#limits

cURL Too many subrequests by single Worker invocation. To configure this limit, refer to https://developers.cloudflare.com/workers/wrangler/configuration/#limits.

cURL Too many subrequests by single Worker invocation. To configure this limit, refer to https://developers.cloudflare.com/workers/wrangler/configuration/#limits

Apply the agreed ingress, insulation, functional or visual acceptance criteria and retain the test record.

Applications and Specimen Integration

IPX9K testing is frequently specified for vehicle lighting, connectors, sensors, control units, battery-related enclosures, motors, outdoor electrical housings, agricultural machinery parts and components exposed to pressure washing. The product category alone does not determine the test: the governing standard and real service risk do.

  • Powered specimens: require protected feed-throughs, suitable internal power and an agreed electrical safety plan.
  • Large or heavy specimens: require a turntable and fixture load review, including eccentric loading.
  • Heat-generating samples: require operating load, duty cycle and heat-release information.
  • Connected components: require cable, coolant, pneumatic or data interfaces that do not compromise spray containment.

Calibration and Acceptance Checks

A defensible test depends on more than the pressure shown on the controller. The acceptance plan should identify which channels are calibrated, the measurement location, allowable tolerance and report format.

Water Conditions

Verify pressure, flow and temperature with instruments suitable for the working range.

Exposure Geometry

Check nozzle angle, specimen distance, alignment and any fixture obstruction.

Motion and Timing

Confirm turntable speed, sequence duration and controller timing.

التوثيق

Retain instrument identification, calibration status, setup photographs, settings and results.

Safety and Maintenance Considerations

High-pressure water at elevated temperature presents burn, injection, electrical and stored-energy hazards. Operators should be trained on interlocks, emergency shutdown, depressurization and safe access. Maintenance should include nozzle inspection, filter cleaning, tank and piping checks, pump inspection, heater scale control, seal inspection and verification of safety devices.

Water quality matters: suspended solids and mineral scale can alter nozzle performance and shorten heater, pump and valve life. Define filtration, water replacement and descaling intervals according to use frequency and local water quality.

Information Needed for an Accurate Quotation

  • Required standard, edition and any customer-specific procedure
  • Specimen dimensions, weight, photographs and mounting orientation
  • Whether the specimen operates during the test and its electrical load
  • Required pressure, flow, temperature, distance, duration and sequence
  • Fixture, turntable and feed-through requirements
  • Site voltage, frequency, water quality, drainage and available floor space
  • Required calibration, acceptance report and third-party verification

الأسئلة المتكررة

What is the difference between IPX9 and IPX9K?

IPX9K is associated with ISO 20653 for road-vehicle electrical equipment, while IPX9 appears in current IEC 60529 provisions for electrical enclosures. They share the concept of high-pressure, high-temperature water jets, but the specified method and edition must be checked before claiming compliance.

Can this chamber also perform IPX5 and IPX6 tests?

Only when compatible IPX5/IPX6 nozzles, water circuits and control functions are included and verified. The IPX9K spray system alone should not be treated as an IPX5/IPX6 system.

Can a specimen be powered and monitored during testing?

Yes, this can be configured when the electrical load, monitoring channels, feed-throughs, isolation and risk controls are defined before manufacture. The safe operating and shutdown procedure must also be agreed.

What water supply does the chamber require?

Purified or appropriately treated water is preferred to reduce deposits and nozzle blockage. Final requirements depend on the specified method, chamber recirculation design and local water quality.

Which parameters should be calibrated?

cURL Too many subrequests by single Worker invocation. To configure this limit, refer to https://developers.cloudflare.com/workers/wrangler/configuration/#limits.

cURL Too many subrequests by single Worker invocation. To configure this limit, refer to https://developers.cloudflare.com/workers/wrangler/configuration/#limits

cURL Too many subrequests by single Worker invocation. To configure this limit, refer to https://developers.cloudflare.com/workers/wrangler/configuration/#limits.

cURL Too many subrequests by single Worker invocation. To configure this limit, refer to https://developers.cloudflare.com/workers/wrangler/configuration/#limits

cURL Too many subrequests by single Worker invocation. To configure this limit, refer to https://developers.cloudflare.com/workers/wrangler/configuration/#limits cURL Too many subrequests by single Worker invocation. To configure this limit, refer to https://developers.cloudflare.com/workers/wrangler/configuration/#limits. cURL Too many subrequests by single Worker invocation. To configure this limit, refer to https://developers.cloudflare.com/workers/wrangler/configuration/#limits.

cURL Too many subrequests by single Worker invocation. To configure this limit, refer to https://developers.cloudflare.com/workers/wrangler/configuration/#limits

cURL Too many subrequests by single Worker invocation. To configure this limit, refer to https://developers.cloudflare.com/workers/wrangler/configuration/#limits.

cURL Too many subrequests by single Worker invocation. To configure this limit, refer to https://developers.cloudflare.com/workers/wrangler/configuration/#limits

فيسبوكلينكدإنXريديتواتساب
معلمات الأبعاد أبعاد الصندوق الداخلي 800×800×800مم
  الأبعاد الإجمالية للآلة 1280×980×1650مم
  قطر الطاولة الدوارة φ400مم
  نطاق ضبط فوهة الرش 100-200مم
  حجم العينة المقترح ≤400×300×300مم
     
معلمات الأداء حركة المرور 14-16 لتر/دقيقة
  نطاق ضغط رش الماء 8000-10000 كيلوباسكال
  سعة التحميل على الطاولة الدوارة 50 كجم
  سرعة الطاولة الدوارة 4-10 دورة/دقيقة
  نطاق درجة حرارة الماء 0-80℃
     
شروط الاستخدام جهد البطارية AC 380V
  الطاقة 10.0 كيلو واط
  حجم المكان المقترح الطول × العرض ≥ 2000×1200 ملم
  ظروف مصدر المياه يفضل استخدام الماء المنقى؛ يمكن توصيل الماء الحنفية خارجيًا.

1. ما هو اختبار IPX9K؟
يستخدم اختبار IPX9K لتقييم مقاومة الماء للمعدات تحت ظروف تدفق الماء عالي الحرارة والضغط العالي.

2. كيف تستخدم غرفة اختبار IPX9K؟
قم بضبط المعلمات ذات الصلة وابدأ تشغيل المعدات، ثم يمكن إجراء اختبار مقاومة الماء.

3. ما هي معايير الاختبار لغرفة اختبار IPX9K؟
تتوافق مع المعيار الدولي لمقاومة الماء IPX9K.

4، كيف يتم تحليل نتائج الاختبار؟
بعد اكتمال الاختبار، سيتم إنشاء سجلات اختبار مفصلة في لوحة التحكم الخاصة بالمعدات. يمكن تصدير هذه السجلات كتقارير للتحليل.

5. ما هي المنتجات المناسبة لغرفة اختبار IPX9K؟
تناسب مجموعة متنوعة من المنتجات مثل المنتجات الإلكترونية، وقطع غيار السيارات، والمعدات الخارجية.

6. ما هي متطلبات الصيانة لغرفة اختبار IPX9K؟
فحص جميع مكونات المعدات بانتظام والحفاظ على نظافتها وفي بيئة تشغيل جيدة.

7. هل يمكن تقديم خطط اختبار مخصصة؟
يمكن تقديم خطط اختبار مخصصة وإعدادات معلمات بناءً على متطلبات العميل.

8، ما هي عملية شراء غرفة اختبار IPX9K؟
يمكنك التواصل مع فريق مبيعات ديروي للحصول على التفاصيل وإكمال عملية الشراء.

 

هاتفهاتف

الهاتف المجاني:
155 8032 7593

واتسابواتساب
واتساب
لينكد إنلينكد إن

ملف لينكد إن الخاص بـ ديروي

رسالةرسالة

اتصل بنا لنقدم لك خدمة حصرية!